Products

 

Analgesic Gas Delivery Systems

Equinox® Advantage Adjustable Analgesic Gas Delivery System

equinox-advantage-front-4-x-3-150-dpi-copyThe O-Two Equinox® Advantage Oxygen/Nitrous Oxide Mixing and Delivery System allows you to control the mix of gas and oxygen a patient is receiving in order to provide the appropriate dose for the patient and the procedure. Designed for use everywhere from the pre-hospital (ambulance), in-hospital (ER, Labor and Delivery etc.) and Dental  (Surgical Procedures), the portable Equinox® Advantage system is easy to use, and eliminates the risk of delivering a hypoxic mixture.

The O-Two Equinox® Advantage  consists of a dual gas (Oxygen and Nitrous Oxide) control module mixing system that maintains the output gas mixture at a constant ratio, over 8 settings, from 25% to 100% oxygen. Each of the two inputs have gas specific built-in alarm systems that generate both visual and audible alarms should either the nitrous oxide or oxygen input fall below 40 PSI. There is also a built-in oxygen fail-safe that will provide 100% oxygen should the nitrous oxide input fall below 35 PSI.

The system has one Rotary Control Knob for turning the unit to the ON or OFF position and one Rotary Control knob for the selection of the Oxygen/Nitrous Oxide mixture. When the unit is turned ON, the output of the O2/N2O gas mixture will only be activated by an inspiratory effort by the patient. The internal demand valve delivers high flow rates with low triggering pressures, reducing the amount of effort required by the patient to maintain the demanded flow. The exhaled gas scavenging port allows the attachment of a scavenging hose to help scavenge exhaled gas away from the patient environment.

The Oxygen Enrichment feature on the O-Two Equinox® Advantage allows the Operator to provide up to 20 L/min of 100% Oxygen through  the Patient Circuit by depressing the O2 Enrichment Button.

Virtually maintenance free, the durable O-Two Equinox® Advantage Analgesic Gas Mixing and Delivery System is a reliable, easy-to-use, safe and efficient way of delivering a range of gas
mixtures for trauma, childbirth or other painful procedures.

Specifications

(Minimum) Peak Flow rate 120 L/Min
Exhalation Resistance 0 to 6 cmH2O @ 60 L/Min
Inhalation Resistance 0 to -6 cmH2O @ 60 L/Min
O2/N2O Mix Settings 25/75, 40/60, 50/50, 60/40,70/30, 75/25, 90/10, 100/0
Input Pressure 50 to 70 PSI (3.5 to 4.8 Bar)
Demand Valve Triggering Pressure -2.0 cmH2O
Pressure Relief Safety Valve 60 cmH2O
Visual and Audible Alarms (@60db) O2 160-200 BPM <40 PSI / N2O 45 – 51 BPM<40 PSI
Operating Temperature 41oF to 104oF (5oC to 40oC)
Storage Temperature -40oF to 140oF (-40oC to 60oC)
Nitrous Oxide Input Connection CGA1040 DISS
Oxygen Input Connection CGA1240 DISS
Patient Valve Dead Space 8 ml
Weight 4.35 lbs (1.96 Kg)
Dimensions WxDxH 9.0” x 7.5” x 4.8” (229 x 190 x 122mm)
Optional Oxygen Enrichment 30 L/min

Part Number and Product Description

01EQ2000 O-Two Equinox® Advantage N2O/O2 Analgesic Gas Mixing and Delivery System c/w O2 and N2O gas supply hoses and single patient use circuit with mouth piece and scavenging hose.
01CV8028-CS Disposable Patient Circuit c/w mouth piece and scavenging hose.   Case of 10
01FV4302-XX 6 foot O2 (Oxygen) Supply Hose with Fittings
01FV4310-XX 6 foot N2O (Nitrous Oxide) Supply Hose with Fittings

 