A/CV is the default mode that appears upon startup of the O-Two e700 ventilator. It offers the widest range of settings and can operate in two ventilation types:
-
Volume-Controlled Ventilation (VCV)
-
Pressure-Controlled Ventilation (PCV)
🔄 How It Works:
-
The ventilator delivers mandatory breaths at a preset rate.
-
Patient effort is not required to initiate breaths (but can trigger them if enabled).
-
If the trigger is disabled, breaths will be delivered purely based on time.
🔧 Default Trigger Sensitivity:
3 L/min (adjustable from 1 to 15 L/min)
📋 Main Settings in A/CV Mode:
Setting | Description |
---|---|
VT (Tidal Volume) | Sets the volume of air delivered per breath (used in VCV only) |
Rate | Number of breaths per minute |
MV (Minute Volume) | Total air delivered per minute (VT × Rate) |
I:E Ratio | Ratio of inspiration to expiration time |
TI (Inspiratory Time) | Duration of the inspiratory phase |
PEEP | Pressure maintained at the end of exhalation |
O₂% | Fraction of inspired oxygen (choice between 60% and 100%) |
MV Max/Min | Alarm limits for high and low minute volume |
P Max/Min | Alarm limits for high and low airway pressure |
🔁 Switching Between Volume and Pressure Control:
-
Volume Mode (default):
-
You set the tidal volume (VT).
-
The ventilator delivers the exact set volume with each breath.
-
-
Pressure Control Ventilation (PCV):
-
Activated by selecting PCV Pressure.
-
VT setting is disabled.
-
The ventilator delivers air until the set pressure is reached, maintained for the preset inspiratory time.
-
Useful when the patient is completely unable to breathe.
-
Cycle is time-triggered and pressure-limited.
-
💡 Summary:
-
A/CV mode is ideal for patients who cannot breathe adequately or at all.
-
Can be tailored for either precise volume delivery or pressure-targeted support.
-
It provides full control over respiratory support, while allowing optional patient triggering.