Multiple Modes, Compact, Lightweight, Easy to Use.
The e700 ventilator ushers in a “new era” in controlled ventilation for resuscitation and patient transport. This electronically controlled, pneumatically powered ventilator provides a range of ventilation solutions to meet the needs of patients across pre-hospital and in-hospital healthcare environments.
The e500 and e700 ventilators usher in a new era in controlled ventilation for resuscitation and patient transport! These electronically controlled, pneumatically powered ventilators provide a range of ventilation solutions for all types of patient requirements and all levels of pre-hospital and in-hospital healthcare professionals.
Simplicity Controllability Functionality Safety Economy
The range of ventilation solutions offered by these devices is unmatched by any other products of their type. With an easily changeable, long-lasting battery pack providing 18–24 hours of operation, the units are self-contained and require only connection to a regulated oxygen supply and a transport ventilator circuit for immediate use.
The easily replaceable battery can be charged while installed in the ventilator or removed for charging and quickly replaced with a fully charged battery pack. The display lighting features both day and night modes, allowing easy visualization of ventilator settings under any ambient lighting conditions.
Designed to accommodate a range of patient sizes, from large adults to infants, both models are compact and lightweight.
The standardized controls on both models provide an ease-of-use concept that is second to none. There are no multiple screens to scroll through to establish patient ventilation parameters. The units feature an initial, preset start-up mode that requires only the selection of patient size to begin ventilation. These products are designed to speed up and simplify even the most complex ventilation setups.
e700 Ventilator Specifications
Power Source:
Compressed Oxygen, 45–87 PSI (3–6 bar)
Circuit Control Source:
Electric
Ventilation modes:
A/C (VCV, PCV), SIMV with PSV, BiLVL with PSV, CPAP with PSV, Mask CPR, and Intubated CPR
Supporting Ventilation:
PSV: 0, 4–35 cmH₂O (±10% or ±2 cmH₂O)
Ventilation Rate:
5–60 BPM (±10% or ±1 BPM)
Minute Volume (L):
Calculated
Tidal Volume (ml):
50 – 2000 (±20ml or ±15%) BTPS*
Tidal Volume in CPR mode (ml):
50 – 1400 (±20ml or ±15%) BTPS *
Maximum Delivered Flow (L/min):
100 – 120
Manually Triggered Ventilation:
Yes. The set flow rate or pressure is delivered for the duration of the inspiratory time, followed by an inspiratory hold.
Maximum Inspiratory hold time:
6 sec.
I:E Ratio:
1:4 – 3:1 (± 20%)
PEEP (cm H2O):
0,4-20 (± 10% or ± 2 cmH2O)
PSV:
OFF, 4 – 35 (± 10% or ± 2 cmH2O)
CPAP (cm H2O):
4-20 (± 10% or ± 2 cmH2O)
O2 (%):
60 or 100 (± 15%)
Pmax (cm H2O):
10 – 80 (± 10% or ± 2 cmH2O)
Pmin (cm H2O):
0 – 20 (± 10% or ± 2 cmH2O)
PCV (cm H2O):
4-50 (± 10% or ± 2 cmH2O)
Ti (sec.):
0.2–9.0 sec. (± 20%)
Trigger sensitivity (L/min):
1–15 L/min, or 2 cmH₂O below baseline in CPAP mode only
Inhalation Pressure (cm H2O):
4-50 (± 10% or ± 2 cmH2O)
Pressure ventilation Termination:
20% – 80% of max. Flow
APNEA back up time (sec.):
10-60 (± 0.5s)
Battery Operating time at room temperature (hrs.):
> 18 hrs for default settings (Data obtained using fully charged new battery)
Altitude compensation:
up to 4000m (13000 feet)
Battery Hot Swap:
No
Built-in Battery charger:
Yes
AC/DC power supply:
100-240 VAC/ 19 VDC, 4.74 A
Patient circuit:
O-Two Electronic Ventilator Circuit
Mounting Bracket:
Mounting brackets for road ambulance and mobile setting
Display:
4.3″ Color TFT
Live monitoring:
Mve,Vte,Paw(AV),PAW(Peak), Rate (bpm), Battery level
Real time waveform:
Pressure or Flow
DAY/NIGHT display mode:
Yes
Parameter settings:
Control Selection Knob
Lock key function:
Yes
Pause function:
Yes
Noise level in normal use:
Less than 65 dBA
Alarms (Visual and Audible):
Gas Supply Pressure, Airway Pressure limits, Minute Volume limits, Battery status, APNEA, Breathing Circuit Integrity, Leakage and Blockage
Audible silence:
Yes, 120 second max
Dimensions (mm)
250 x 200 x 155
Weight (Kg/Lbs) with/without Battery
2.4/1.77 | 5.29/3.9
Internal Volume of the complete respiratory system (reusable and disposable):
approx. 690 ml without mask | approx. 800 ml with mask
Dead space of patient valve with elbow:
Approx. 35 ml
Compliance (disposable) hose system:
16.6 ml/kPa
Resistance of Patient Hose system (Inhalation and Exhalation):
Less than 6 cmH2O at 60 l/min & Less than 6 cmH2O at 30 l/min
Environment condition
Ventilator:
Operating Continuous – 18oC to +40o C, Rh: 15% to 95% | OperatingTransient** – 20o C to +50o C, Rh: 15% to 95% | Storage – 40o C to +60o C, Rh: 15% to 95%
Battery Pack:
Charge: 0oC to +40oC | Discharge | – 20oC to +60oC | Storage: – 20oC to +35oC, low humidity and no corrosive gas atmosphere.
Patient Circuit:
Operating: – – 18oC to +50oC, Rh: 15% to 95% | Storage: – – 20oC to +60oC, Rh: 15% to 95%
* BTPS: Volume measurements corrected to Body temperature 37oC and Barometric pressure 101.3Kpa under saturated conditions (100% Humidity). Note: Measurement uncertainty: 5% for volume parameters and 6% for pressure parameters.
This product has a two year warranty against manufacturers defects.
Device class per MDD:
II b
Classification According to IEC 60601-1
Protection Against Electric Shock: Class II, Type BF
Protection Against Water: IPX4
01EVE700
e700 – Electronic Automatic Transport Ventilator, complete with Disposable Patient Circuit, Resuscitation Mask, 6-Foot Supply Hose with 9/16″ DISS Fittings, Calibrated Test Lung, and Power Supply (Specify Country of Use)
Each
Accessories
01CV8030-CS
O-Two Medical Single-Use Electronic Transport Ventilator 6-Foot Circuit, with Protective Sleeve. Compatible with the e500 and e700
Case/10
01CV7035
O-Two Medical eSeries Ventilator Smart Mount – Multi-Configuration Mounting Bracket (Ambulance Cot, Hospital Stretcher, Bed, Roll Stand)
Each
01CV8040-CS
O-Two Medical eSeries Ventilator Replacement Intake Filter/Cover
Case/10
01TA1852
O-Two Medical eSeries Ventilator Replacement 1-Litre Test Lung with Compliance Restrictor
Each
01FV4303-DISS
O-Two 6 Foot (1.85 Meter) O2 Supply Hose with 9/16 DISS Nut and 9/16” DISS Nut Ventilator Connection
Each
01CV9100
O-Two Medical eSeries Replacement Lithium-Ion Battery
Each
01CV0106
Power Supply Cord for eSeries Power Supply (US/CA)
Each
01CV0102-EU
Power Supply Cord for eSeries Power Supply (European)
Each
02FM4999-CS
O-Two Universal Mask
Case/12
01CV7050
eSeries Automatic Transport Ventilator Carrying Case with Sling-Style Shoulder Strap – Specifically Designed for eSeries
Each
02FM5207-CS
O-Two Disposable Cuffed Child/Small Adult #3 Mask with Hook Ring
Case/12
02FM5000-CS
O-Two Disposable Cuffed Adult #4 Mask with Hook Ring
Case/12
02FM5001-CS
O-Two Disposable Cuffed Large Adult #5 Mask with Hook Ring
Case/12