About Lesson
To modify any parameter in the chosen mode:
-
Rotate the control knob to place the yellow cursor on the desired setting.
-
Press the knob to highlight the parameter.
-
Rotate again to adjust the value.
-
Press once to set the value.
-
Press twice to confirm the new setting.
• If no confirmation is made within 10 seconds, the ventilator will revert to previous values automatically.
When the ventilation mode is changed, Section 6 of the screen layout will update to show the relevant setup parameters for the new mode.