Equinox Relieve

Designed for use in the pre-hospital (ambulance) and in-hospital use (ER, Labor and Delivery etc.), the O-Two Equinox® Relieve is intended to provide a 50/50% mixture of nitrous oxide and oxygen, on demand, to a conscious, spontaneously breathing, patient for the relief of pain due to trauma or other conditions where inhalation analgesia is clinically specified. The system has no operator or patient adjustable controls eliminating the risk of delivering a hypoxic mixture. It only requires the attachment to pressure regulated medical gas oxygen and nitrous oxide cylinders and the patient circuit with mouth piece and scavenging hose to be ready for use.

The O-Two Equinox® Relieve consists of a dual gas (Oxygen and Nitrous Oxide) control module mixing system that maintains the output gas mixture at a constant 50/50 mix. Each of the two inputs have gas specific built-in alarm systems will generate both visual and audible alarms should either the nitrous oxide or oxygen input fall below 40 PSI, and the device will be automatically shut off should either the nitrous oxide or oxygen input fall below 35 PSI.

The O-Two Equinox® Relieve is also equipped with a secondary “fail safe” circuit that will activate an alarm and shut off the device should internal malfunction occur in the mixer or any internal hoses rupture or kink.

O-Two Equinox® Relieve N2O/O2 Analgesic Gas Mixing and Delivery System has only one control for turning ON or OFF the device. When it is turned ON, the output of the N2O/O2 gas mixture will only be activated by an inspiratory effort by the patient. The internal demand valve delivers high flow rates with low triggering pressures, reducing the amount of effort required by the patient to maintain the demanded flow. During use, the scavenging hose on the circuit assists in scavenging the patient’s exhaled air away from the patient environment.

Virtually maintenance free, the durable O-Two Equinox® Relieve N2O/O2 Analgesic Gas Mixing and Delivery System is a simple, safe and efficient way of delivering pain relief during trauma, childbirth or other painful procedures.

The optional Oxygen Enrichment feature allows the rescuer flush the system by depressing the manual button.

Minimum Flow Rate

120 L/min

Exhalation Resistance

0 to 6 cmH2O @ 60 L/Min

Inhalation Resistance

0 to -6 cmH2O @ 60 L/Min

Oxygen Concentrations

42.5 – 57.5%y

Oxygen Enrichment Flow (01eq100f Only)

30L/min

Input Pressure

50 to 70 PSI (3.5 to 4.8 Bar)

Low Input Pressure Alarm

40 +- 1 PSI (2.7 Bar)

Device Cut Off Pressure

35 ± 4 PSI (2.7 Bar)

Demand Valve Triggering Pressure

-1.5 to -2.5 cmH2O

Device Input Connections

Nitrous Oxide: CGA1040 DISS Oxygen: CGA1240 DISS

Patient Connector

15/22 mm

Patient Circuit (Single Use)

Dual limb circuit with one-way valves and 19mm scavenging Port Connection on the exhalation limb.

Patient Valve Dead Space

32 ml

Operating Temperature

41o F to 104o F (50 C to 40o C)

Storage Temperature

-40o F to 140o F (-400 C to 600 C)

Weight

3.3 lbs (1.5 Kg)

Failsafe protection

~ 2 PSI diff erential pressure between O2 and N2O output gas. Prevents the output of high nitrous oxide concentrations

Dimensions Wxdxh

8.9” x 6.6” x 3.9” (226 x 168 x 99mm)

Equinox® Relieve Specifications

This product has a two year warranty against manufacturer defects.

Equinox® Relieve Ordering Information

01EQ1000

Equinox® Relieve N2O/O2 Analgesic Gas Mixing and Delivery System c/w O2* and N2O** Supply Hoses, Single-Use Patient Circuit with Scavenger Hose and Adapter with Mouthpiece

Each

01EQ1000F

Equinox® Relieve N2O/O2 Analgesic Gas Mixing and Delivery System with Oxygen Enrichment Button c/w O2* and N2O** Supply Hoses, Single-Use Patient Circuit with Scavenger Hose and Adapter with Mouthpiece

Each

01CV8028-CS

Single-Use 6 Foot Circuit with Scavenger Hose and Adapter with Mouthpiece

Case/10

02RT1303

Disposable Mouth Piece (Individually Wrapped)

Case/50

17MP1997

O-Two 90o (Male to Male) adjustable Oxygen (O2) DISS elbow assy

Each

17MP1998

O-Two 90o (Male to Male) adjustable Nitrous Oxide (N2O) DISS elbow assy

Each

Specification Sheet

User Manual

Study Guide

ACCESSORIES