In this part of the O-Two e700 ventilator training video, viewers are guided through the proper steps to connect the patient breathing circuit and gas supply:
-
Oxygen Supply Connection:
-
Attach the high-pressure oxygen supply hose to the 9/16-inch DISS (Diameter Index Safety System) connector.
-
Hand-tighten only — do not use tools.
-
-
Sensor Hose Setup:
-
Connect the sensor hose with the lower locking mechanism to Sensor Connector #1.
-
Attach the second sensor hose to Connector #2 by simply pushing it on.
-
-
Breathing Circuit Connection:
-
Connect the 22mm breathing circuit hose to the male 22mm outlet on the ventilator.
-
The other end of the oxygen supply hose should be connected to a regulated oxygen source — either a gas cylinder or a wall outlet.
-
-
Gas Supply Activation:
-
Turn on the gas supply slowly and fully.
-
-
Patient Safety Precaution:
-
Do not connect the breathing circuit to the patient while attaching the circuit to the ventilator.
-
-
Optional Filter Use:
-
A breathing circuit filter may be used, but it must be placed between the patient connection and the mask or ET tube.
-
-
Elbow Connector Note:
-
Ensure the elbow connector is always present and properly connected, as it is prone to falling off.
-
This connector links the breathing circuit to the patient’s mask or endotracheal (ET) tube.
-
This segment emphasizes accuracy and safety, ensuring that each component is securely and correctly connected before initiating ventilation.