This part of the video explains how the O-Two e700 ventilator is powered and provides step-by-step instructions for installing the internal battery.
The O2E700 can operate using either its internal rechargeable battery or an external AC/DC power supply. For safety and uninterrupted ventilation, a fully charged battery must always be installed, even when using an external power source. This ensures continuous operation in case of power failure or disconnection.
Battery Installation Steps:
-
Ensure safety first – Turn off the ventilator and unplug it from the electrical supply.
-
Open the battery compartment by turning the yellow screw knobs on the cover counterclockwise.
-
Connect the battery leads carefully.
-
Close the battery compartment and turn the yellow knobs clockwise to secure it.
• Important Tip:
When replacing the battery, never pull it by the connectors. Instead, use the battery standoff to gently remove it after disconnecting the leads.