This section of the video provides an overview of the user interface and control elements used to operate the O-Two e700 ventilator. The device is managed using a combination of membrane keys and a control selection knob, allowing users to access and adjust ventilation settings easily.
Main Controls and Their Functions:
-
On/Off Button
-
Press for 1 second to power on the ventilator.
-
Press for 4 seconds to power off the ventilator.
-
-
Control Selection Knob
-
Rotate to navigate settings and modes.
-
Press to confirm selections.
-
-
Manual Key
-
During exhalation, pressing this key delivers a mandatory breath using the set flow or pressure.
-
Holding the key after inspiratory time triggers the inspiratory hold function (up to 6 seconds), maintaining lung pressure by closing the exhalation port.
-
After the hold time, the ventilator automatically transitions to the exhalation phase.
-
Enabled in all modes except CPAP and CPR.
-
-
Brightness Key
-
Adjusts screen color and contrast for better visibility.
-
-
Silence Key
-
Mutes audible alarms for 2 minutes.
-
Can be used proactively even when no alarm is active.
-
Displays a mute symbol on the screen.
-
-
Pause Key
-
Pauses ventilation while keeping other controls active (except Manual).
-
Activates a pause symbol and confirmation prompt on the screen.
-
If not confirmed by pressing the knob within 10 seconds, the pause is canceled.
-
-
Lock Key
-
Disables all buttons except On/Off, Silence, Brightness, and the Control Knob.
-
Displays a lock icon to indicate the function is active.
-
Press again to unlock.
-
-
Waveform Selection Key
-
Toggles between pressure and flow waveforms on the display.
-
-
Cancel Key
-
Cancels the last unconfirmed changes and restores previous settings.
-